STUDY OF 3D PRINTING HIPS BREAST PHANTOM MATERIAL: EFFECT OF EXPOSURE VOLTAGE VARIATION AND NODULE SIZE
DOI:
https://doi.org/10.22437/jiituj.v9i3.42846Keywords:
3D Printer, Breast Phantom, CT-Number, CT-Scan, HIPS, PLAAbstract
A study has examined the characteristics of 3D printer materials of HIPS and PLA. The study began by printing HIPS materials that resembled breast organs and PLA materials that resembled breast cancer nodules. Then it exposed them using a CT scan with voltage variations of 80 kV, 100 kV, and 120 kV. The parameters studied were compared with the reference breast organ. The results obtained at these voltages obtained CT-number values sequentially (-85.14+0.25) ; (-64.79+0.21) HU; (-55.08+0.18) HU values from the reference 35 to -94 HU, relative electron density values sequentially 0.915; 0.935; 0.945 with a reference value of 0.950, the density value of the material obtained is (0.921+0.002) g/, The EDG value is 3.24 x from ICRU 1989 was 3.33 x, The EDV value is (3.00 + 0.01) x from ICRU report 44 was 3.18 x 10, the value is 3.50 the value of ICRU data is 3.28, and the effective dose values are respectively 2.53 mSv, 2.83 mSv, 3.20 mSv. Based on these characteristics, the results indicate that the breast phantom matches the characteristics of the breast organ, allowing it to be used to predict the presence of breast cancer, as it shares the same characteristics. The results of exposure to the lung cancer nodule phantom on the CT-Scan image of the 2 mm nodule size are not visible. The results of this study can be used as a consideration for radiologists when choosing a voltage based on the predicted size of the breast cancer nodule.
Downloads
References
Ali, A., Wahab, R., Huynh, J., Wake, N., & Mahoney, M. (2020). Imaging properties of 3D printed breast phantoms for lesion localization and Core needle biopsy training. 3D Printing in Medicine, 6, 1-6. https://doi.org/10.1186/s41205-020-00058-5
Almalki, H., Marashdeh, M. W., Alsuhybani, M., & Almurayshid, M. (2024). Evaluation of radiation attenuation properties of breast equivalent phantom material made of gelatin material in mammographic X-ray energies. Journal of Radiation Research and Applied Sciences, 17(1), 100817. https://doi.org/10.1016/j.jrras.2024.100817
Anwar, Y., Zarzani, TR, & Chermanto. (2023). Legal responsibility of medical physicists for quality of radiology operational permits at bhayangkara hospital, Banda Aceh. Sibatik Journal , 2(8), 2263-2282 . https://doi.org/10.54443/sibatik.v2i8.1196
Belarra, A., Chevalier, M., Garayoa, J., Hernández-Girón, I., & Valverde, J. (2024, May). Measuring effective attenuation coefficients of 3D printing materials for anthropomorphic breast phantoms. In 17th International Workshop on Breast Imaging (IWBI 2024) (Vol. 13174, pp. 352-360). SPIE. https://doi.org/10.1117/12.3027031
Byng, J. W., Mainprize, J. G., & Yaffe, M. J. (1998). X-ray characterization of breast phantom materials. Physics in Medicine & Biology, 43(5), 1367. https://doi.org/10.1088/0031-9155/43/5/026
Carton, A. K., Bakic, P., Ullberg, C., Derand, H., & Maidment, A. D. (2011). Development of a physical 3D anthropomorphic breast phantom. Medical physics, 38(2), 891-896. https://doi.org/10.1118/1.3533896
Cuong, NLQ, Minh, NH, Cuong, HM, Quoc, PN, Van Anh, NH, & Van Hieu, N. (2018). Porosity estimation from High-Resolution CT Scan images of rock samples using hounsfield units. Open Journal of Geology, 8(10), 101- 109. https://doi.org/10.4236/ojg.2018.810061
Delombaerde, L., Petillion, S., Weltens, C., De Roover, R., Reynders, T., & Depuydt, T. (2020). Development of 3D‐printed breast phantoms for end‐to‐end testing of whole breast volumetric arc radiotherapy. Journal of Applied Clinical Medical Physics, 21(8), 315-320. https://doi.org/10.1002/acm2.12976
di Franco, F., Mettivier, G., Sarno, A., Varallo, A., & Russo, P. (2019). Manufacturing of physical breast phantoms with 3D printing technology for X-ray breast imaging. In 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) (pp. 1-5). IEEE https://doi.org/10.1109/NSS/MIC42101.2019.9059986
Di Meo, S., Cannatà, A., Morganti, S., Matrone, G., & Pasian, M. (2022). On the dielectric and mechanical characterization of tissue‐mimicking breast phantoms. Physics in Medicine & Biology, 67(15), 155018. https://doi.org/10.1088/1361-6560/ac7bcc
Dukov, N., Bliznakova, K., Teneva, T., Marinov, S., Bakic, P., Bosmans, H., & Bliznakov, Z. (2021). Experimental evaluation of physical breast phantoms for 2D and 3D breast x-ray imaging techniques. In 8th European Medical and Biological Engineering Conference: Proceedings of the EMBEC 2020, November 29–December 3, 2020 Portorož, Slovenia (pp. 544-552). Springer International Publishing. https://doi.org/10.1007/978-3-030-64610-3_62
Endra, K., & Villaflor, G. M. (2024). Integration of the POE model and metaphoral thinking in student worksheets: Improving mathematical reasoning abilities in the modern education era. Journal of Educational Technology and Learning Creativity, 2(1), 41-53. https://doi.org/10.37251/jetlc.v2i1.981.
Esposito, G., Mettivier, G., Bliznakova, K., Bliznakov, Z., Bosmans, H., Bravin, A., & Russo, P. (2019). Investigation of the refractive index decrement of 3D printing materials for manufacturing breast phantoms for phase contrast imaging. Physics in Medicine & Biology, 64(7), 075008. https://doi.org/10.1088/1361-6560/ab0670
Fiarka, F., Oktamuliani, S., & Nuraeni, N. (2023). Effect of voltage on CTDIvol Value, dose length product and effective dose in computed tomography (CT) abdomen examination: Study at Dr. M. Djamil central general hospital. Journal of Physics Unand, 12(4), 591-597. https://doi.org/10.25077/jfu.12.4.591-597.2023
Globocan. (2022). Cancer Incident in Indonesia. International Agency for Research on Cancer, 858: 1–2.
Gómez, Á. ML, & Mourão, A.P. (2022). Study of materials for the construction of phantoms for radiodiagnostic tests. In Proceedings of the 5. SENCIR: Brazilian National Week on Nuclear and Energy Engineering and Radiation Sciences, 1–9. https://doi.org/10.37885/221110974
Gounhalli, S. G., Basavaraj, S., & Hanagodimath, S. M. (2022). Mass attenuation coefficients, effective atomic number, electron density, and kerma of BBPD and BBPO Organic Molecules. Journal of the Maharaja Sayajirao University of Baroda , 56(2), 9-18.
Gusriani, Umami, N., Noviyanti, NI, Rusmiati, & Fitri, G. (2023). Early detection of breast cancer with clinical breast examination (SADARI). Borneo Community Service Journal, 7(1), 51-55. https://doi.org/10.35334/jpmb.v7i1.3326
Grigorova, N., Kaloyanova, D., Dukov, N., & Bliznakova, K. (2024, November). Ultrasound breast phantom for breast biopsy training. In 2024 E-Health and Bioengineering Conference (EHB) ( 1-4). IEEE. https://doi.org/10.1109/EHB64556.2024.10805633
Guswantoro, T., Supratman, AS, & Asih, IS (2020). Characterization of alginate as a soft tissue equivalent material for radiotherapy. EduMatSains: Journal of Education, Mathematics and Science, 4(2), 125-138. https://doi.org/10.33541/edumatsains.v4i2.1378
Habibi, M. W., Jiyane, L., & Özşen, Z. (2024). Learning revolution: The positive impact of computer simulations on science achievement in madrasah ibtidaiyah. Journal of Educational Technology and Learning Creativity, 2(1), 13-19. https://doi.org/10.37251/jetlc.v2i1.976.
Hatamikia, S., Kronreif, G., Unger, A., Oberoi, G., Attorney, L., Unger, E., & Lorenz, A. (2022). 3D Printed Patient-Specific thorax phantom with realistic heterogeneous bone radiopacity using filament printer technology. Zeitschrift für Medizinische Physik, 1(1), 1-15. https://doi.org/10.1016/j.zemedi.2022.02.001
Halimah, H., Putri , D. E., Wulandari, W., Adewumi, S. E., & Arce-Calderón, X. (2024). Contextual pop up book as an innovative learning media in social science subjects in elementary schools. Journal of Educational Technology and Learning Creativity, 2(2), 209-216. https://doi.org/10.37251/jetlc.v2i2.1121.
Hermanto, A., Aridamayanti, BG, Iswahyudi, UA (2023). The Relationship between Active and Passive Smokers and Knowledge of Lung Cancer in Students at SMAN 1 Wongsorejo. Profesional Health Journal, 4(2): 281-286. http://dx.doi.org/10.54832/phj.v4i2.383
He, Y., Liu, Y., Dyer, B. A., Boone, J. M., Liu, S., Chen, T., & Qiu, J. (2019). 3D-printed breast phantom for multi-purpose and multi-modality imaging. Quantitative imaging in medicine and surgery, 9(1), 63. https://doi.org/10.21037/qims.2019.01.05
Irsal, M., Nurbaiti, Mukhtar, AN, Gunawati, S., & Hidayat, W. (2021). The Effect of Tube Voltage on Image Quality in Head Computed Tomography Examination using Iterative Reconstruction. Journal of Physics Theory and Applications, 9(1):103-110. http://dx.doi.org/10.23960/jtaf.v9i1.2715
Jarnawi, M., Haeruddin, H., Werdhiana, I. K., Syamsuriwal, S., & Mu’aziyah, S. E. S. (2025). Integrating thinking styles into differentiated instruction: enhancing learning outcomes in science education. Integrated Science Education Journal, 6(1), 47-53. https://doi.org/10.37251/isej.v6i1.1328
Katsikari, K., Francis, N., Maalej, N., Pitsalidis, C., & Raja, A. Y. (2024). 3D-Printed anthropomorphic phantom framework for spectral photon-counting CT: A breast phantom application. IEEE Access. https://doi.org/10.1109/ACCESS.2024.3521476
Keenan, K. E., Wilmes, L. J., Aliu, S. O., Newitt, D. C., Jones, E. F., Boss, M. A., ... & Hylton, N. M. (2016). Design of a breast phantom for quantitative MRI. Journal of Magnetic Resonance Imaging, 44(3), 610-619. https://doi.org/10.1002/jmri.25214
Khoirot, RM, Hakim, MH, Alam, Y., & Sekar, R. (2023). Analysis of DRL values of esak/inak parameters of thorax AP/PA X-Ray canon examination. Journal of Science Nusantara, 3(2): 61-68. https://doi.org/10.28926/jsnu.v3i2.1041
Kunert , P. , Schlattl, H. , Trinkl .S , Glussani, A., Klein, L., Janich, M., Reichert, D., & Brix, G. (2023). Reproduction of a conventional anthropomorphic female chest phantom by 3D-Printing: Comparison of image contrasts and absorbed doses in ct. Med Phys, 50, 4734 – 4743. https://doi.org/10.1002/mp.16587
Laksono, P. J., Suhadi, S., & Efriani, A. (2025). Unveiling STEM education conceptions: Insights from pre-service mathematics and science teachers. Integrated Science Education Journal, 6(1), 54-61. https://doi.org/10.37251/isej.v6i1.1387
Lee, J. S., Jo, Y. I., Kang, Y. R., Kye, Y. U., Il, P., & Lee, D. Y. (2020). Filament material evaluation for breast phantom fabrication using three-dimensional printing. Nuclear Technology and Radiation Protection, 35(4), 372-379. https://doi.org/10.2298/NTRP2004372L
Lee, D. Y., Jo, Y. I., & Yang, S. H. (2021). Development of breast phantoms using a 3D printer and glandular dose evaluation. Journal of Applied Clinical Medical Physics, 22(10), 270-277. https://doi.org/10.1002/acm2.13408
Leonov, D., Venidiktova, D., Costa-Júnior, J. F. S., Nasibullina, A., Tarasova, O., Pashinceva, K., & Saikia, M. J. (2024). Development of an anatomical breast phantom from polyvinyl chloride plastisol with lesions of various shape, elasticity and echogenicity for teaching ultrasound examination. International Journal of Computer Assisted Radiology and Surgery, 19(1), 151-161. https://doi.org/10.1007/s11548-023-02911-4
Listiyani, I., Nismayanti, A., Maskur, Kasman, Ulum, MS, Rahman, A., & Samad, S. (2021). Noise Level Analysis of CT-Scan Image Results on Head Phantom with Variations in Tube Voltage and Slice Thickness. Gravitation , 20(1), 5–9. http://dx.doi.org/10.22487/gravitasi.v20i1.15517
Mettivier, G., Sarno, A., Varallo, A., & Russo, P. (2022). Attenuation coefficient in the energy range 14–36 keV of 3D printing materials for physical breast phantoms. Physics in Medicine & Biology, 67(17), https://doi.org/175012. 10.1088/1361-6560/ac8966
Malliori, A., Daskalaki, A., Dermitzakis, A., & Pallikarakis, N. (2020). Development of physical breast phantoms for X-ray imaging employing 3D printing techniques. The Open Medical Imaging Journal, 12(1). http://dx.doi.org/10.2174/1874347102012010001
Marashdeh, M., & Abdulkarim, M. (2023). Determination of the attenuation coefficients of epoxy resin with carbopol polymer as a breast phantom material at low photon energy range. Polymers, 15(12), 2645. https://doi.org/10.3390/polym15122645
Martinez, L.C., Calzado, A., Rodriguez, C., Gilarranz, R., & Manzanas, M.J.(2012). A parametrization of the CT number of a substance and its use for stoichiometric calibration. Physica Medica, 28(1), 33-42. https://doi.org/10.1016/j.ejmp.2011.02.001
Ock, J., Lee, S., Kim, T., Hong, D., Kim, M., Ko, B. S., & Kim, N. (2021). Accuracy evaluation of a 3D printing surgical guide for breast-conserving surgery using a realistic breast phantom. Computers in Biology and Medicine, 137, 104784. https://doi.org/10.1016/j.compbiomed.2021.104784
Plummer, M., de Martel, C., Vignat, J., Ferlay, J., Bray, F., & Franceschi, S. (2016). Global burden of cancers attributable to infections in 2012: a synthetic analysis. The Lancet Global Health, 4(9), e609-e616. https://doi.org/10.1016/s2214-109x(16)30143-7
Pohan, M. Y., Siregar, T. Z., & Panjaitan, B. (2022). Analisa paparan radiasi pada instalasi radiologi di rumah sakit islam malahayati medan tahun 2021 [Analysis of radiation exposure in the radiology installation at the Malangayati Islamic Hospital, Medan in 2021]. Jurnal Ilmiah Ilmu Terapan Universitas Jambi, 6(1), 66–72. https://doi.org/10.22437/jiituj.v6i1.19333
Pratiwi, RF, Pulungan, ES, & Andini, D. (2023). The effect of exposure factors on the quality of radiographic images in thorax examination. Indonesian Radiographer Journal, 6(1), 38-41. https://doi.org/10.55451/jri.v6i1.173
Pritchard, P.E. (1992). Studies on the bread-improving mechanism of fungal alpha-amylase. Journal of Biological Education, 26(1), 14–17. https://doi.org/10.1037/0012-1649.44.4.1055.
Rachmanto, T. B., & Akande, I. O. (2024). Utilization of information technology in increasing the effectiveness of citizenship learning. Journal of Educational Technology and Learning Creativity, 2(2), 217-222. https://doi.org/10.37251/jetlc.v2i2.1140.
Rifda, DZ, Shaluhiyah, Z., & Surjoputro. (2023). Phenomenological study of breast cancer patients in an effort to improve quality of life: Literature review. Indonesian Health Promotion Publication Media, 6(8): 1495-1500. https://doi.org/10.56338/mppki.v6i8.3513
Riza, EI, Budiyantoro, C., & Nugroho, AW (2020). Increasing flexural strength of 3D printing products made of PETG with optimization of process parameters using the taguchi method. Media Mesin: Jurnal Teknik Mesin , 21(2), 66-75. https://doi.org/10.23917/mesin.v21i2.10856
Sindi, R., Wong, Y. H., Yeong, C. H., & Sun, Z. (2020). Development of patient-specific 3D-printed breast phantom using silicone and peanut oils for magnetic resonance imaging. Quantitative imaging in medicine and surgery, 10(6), 1237. https://doi.org/10.21037/qims-20-251
Surahmi, N., Sari, K., Ulliana, U., & Badari, M. (2023). Level of accuracy of intraoral examination in impaction cases using general purpose devices in radiology installation. Journal of Health Research, 15(2), 246-252. https://doi.org/10.34011/juriskesbdg.v15i2.2134
Spytska, L. (2025). Digital technology and mental health: Unveiling the psychological impact of modern digital habits. Jurnal Ilmiah Ilmu Terapan Universitas Jambi, 9(1), 348–365. https://doi.org/10.22437/jiituj.v9i1.38238
Schopphoven, S., Cavael, P., Bock, K., Fiebich, M., & Mäder, U. (2019). Breast phantoms for 2D digital mammography with realistic anatomical structures and attenuation characteristics based on clinical images using 3D printing. Physics in Medicine & Biology, 64(21), 215005. https://doi.org/10.1088/1361-6560/ab3f6a
Utami, N. W., & Saptiari, N. N. (2020). Penerapan data mining untuk klasifikasi penyebab kematian menggunakan algoritma support vector machine. Jurnal Ilmiah Ilmu Terapan Universitas Jambi, 4(2), 234–240. https://doi.org/10.22437/jiituj.v4i2.13268
Widyanengsih, E., Fitria, LN, Jabar, MA, Dzulfiana, N., A, SR, Cahyati, SAW, Salsabila, T., & Yuniarsih, N. (2023). Review article: Nanoparticle drug delivery system in cancer cells. Journal of Education and Counselling, 5(1), 1133–1138. https://doi.org/10.31004/jpdk.v5i1.11122
Yang, K., Burkett, G., & Boone, J. M. (2014). A breast-specific, negligible-dose scatter correction technique for dedicated cone-beam breast CT: A Physics-Based approach to improve Hounsfield Unit Accuracy. Physics In Medicine & Biology, 59(21), 487–505. https://doi.org/10.1088/0031-9155/59/21/6487
Yunianto, M., & Ranny, LN (2025). Study of EPC type 3D printer materials as a kidney organ phantom with variations in CT-Scan exposure voltage and kidney cancer nodule size. In Journal of Physics: Conference Series (Vol. 2945, No. 1, p. 012027). IOP Publishing. https://doi.org/10.1088/1742-6596/2945/1/012027
Yunianto, M., Cari, C., & Sari, APK (2024). Analysis of the effect of changes in thickness and infill density of thermoplastic polyurethane and polyethylene terephthalate glycol materials as a Radiological Phantom. Aceh International Journal of Science and Technology, 13(2). https://doi.org/10.13170/aijst.13.2.40201
Yunianto, M., Yani, S., Anwar, F., Suparmi, A., Cari, C., Sarwidi, S., & Ardyanto, TD (2024). Comparison of CT number and electron density in 3D printing materials for phantom applications. In AIP Conference Proceedings (Vol. 3074, No. 1). AIP Publishing. https://doi.org/10.1063/5.0211295
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2025 Mohtar Yunianto, Nanda Astika Devi, Fuad Anwar, Adnan Cindyana

This work is licensed under a Creative Commons Attribution 4.0 International License.











