Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
Article  
Verification of The Method for Determining Calcium Content in  
Animal Feed  
Syafrinal1* , Hafnimardiyanti1 , Selfa Dewati Samah1 , Pevi Riani1 , Renny Futeri1  
Nurmaliza1  
,
1Chemical Analysis, ATI Padang Polytechnic Jl. Bungo Pasang Tabing, Padang, Sumatera Barat,  
25171, Indonesia  
Abstract  
Calcium is one of the essential macromineral elements in animal feed. To ensure that animal feed meets  
nutritional and safety standards, accurate and validated analytical methods are needed to determine  
calcium levels, specifically using the ICP-OES (Inductively Coupled Plasma Optical Emission Spectroscopy)  
technique, which is known for its high sensitivity, enabling the rapid and precise detection of calcium  
elements at low concentrations. This study aims to verify the method for analyzing calcium content in  
animal feed, referring to method verification parameters, including linearity, accuracy, precision  
(repeatability test), limit of detection (LoD), method detection limit (MDL), and limit of Quantitation (LOQ).  
The verification process was conducted using calcium standard solutions at various concentrations and  
analysis of animal feed samples according to SNI 3148.2:2009 using Inductively Coupled Plasma (ICP-  
OES) with a linearity test parameter coefficient of correlation (r) value of 0.9991, precision with a % RSD  
value of 2.18%, accuracy (recovery) reaches 90-99%, with a detection limit (LoD) of 0.0745 mg/L and a  
Limit of Quantitation (LoQ) of 0.9060 mg/L, all of which meet the acceptance criteria set by AOAC. These  
results prove that the ICP-OES technique is suitable for use as a test method for determining the calcium  
content in animal feed, can be adopted as a standardized routine procedure.  
Keywords: Animal Feed, calcium, ICP-OES, method verification  
*
Corresponding author  
Email address: rinal1450@gmail.com  
Received August 29th 2025; Accepted December 01st 2025; Available online December 25th 2025  
Copyright © 2025 by Authors, Published by Chempublish Journal. This is an open access article under the CC BY License  
313  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
Graphical Abstract  
Introduction  
reproducible [5]. One of the analytical  
methods currently relied upon is Inductively  
Animal feed is one of the most important  
components in modern livestock production  
Coupled  
Plasma  
Optical  
Emission  
Spectroscopy (ICP-OES). This technology  
allows for the simultaneous detection of  
minerals with high sensitivity and low  
detection limits, making it highly suitable for  
complex matrices such as animal feed [6].  
However, the use of the ICP-OES method for  
determining mineral elements must first  
systems  
[1].  
Global  
animal  
protein  
consumption increased by approximately  
14% over the past two decades. This  
underscores the need for feed quality  
assurance,  
including  
essential  
nutrient  
content such as calcium (Ca) [2]. Calcium is  
one of the essential macromineral elements  
in animal feed, playing a primary role in bone  
undergo  
a
verification stage to meet  
analytical  
validity standards. Method  
formation,  
transmission, and blood coagulation [3].  
Calcium requirements vary greatly  
muscle  
contraction,  
nerve  
verification is an important process in  
ensuring that the method used meets  
performance  
accuracy, precision,  
specifications  
linearity,  
such  
limit  
as  
of  
depending on the livestock species, age, and  
physiological stage. Therefore, accurate  
determination of calcium levels is crucial in  
feed formulation [4].  
detection, and limit of quantitation [7].  
International regulatory bodies such as  
AOAC and SNI have emphasized the  
importance of analytical method verification  
processes in ensuring the quality of  
laboratory test result.  
In practice, the analysis of mineral content  
such as calcium in feed materials requires  
methods that are not only sensitive and  
accurate, but also precise, selective, and  
314  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
Various previous studies have developed  
methods to measure calcium in various food  
materials using ICP-OES, such as in milk [8],  
nuts [9], persian tahini (ardeh) ([10] and  
sorghum [11] but there are no publications  
that can be detected and measured with a  
certain degree of confidence and the  
method detection limit (MDL) to determine  
the actual detection limit of the entire  
analytical method [16].  
that  
comprehensively  
examine  
the  
This research also contributes scientifically  
by presenting empirical data that fills a gap  
in the literature regarding the application  
and validation of ICP-OES in livestock feed  
matrices. Additionally, the results of this  
study will serve as a reference for testing  
laboratories, certification bodies, and feed  
manufacturers in establishing accountable  
verification of this method's performance  
parameters in the context of animal feed.  
Some other studies are still limited to  
determining the general levels of mineral  
elements  
verification aspects of the methods used.  
This research gap underscores the  
importance of conducting a study to verify  
the ICP-OES method for determining Ca  
levels in animal feed. Therefore, verifying  
this method is of significant value both  
scientifically and practically, especially in  
supporting the livestock feed industry, which  
relies on accurate and reliable analytical data  
[12].  
without  
emphasizing  
the  
mineral  
testing  
protocols.  
With  
this  
background, the main objective of this  
research is to verify the method for  
determining the calcium (Ca) content in  
animal feed using ICP-OES based on the  
parameters of linearity, limit of linearity,  
precision, accuracy, limit of detection,  
method  
Quantitation,  
detection  
as  
limit,  
regulated  
and limit  
in  
of  
SNI  
This research not only determines the Ca  
content using the ICP-OES technique but  
also conducts a thorough verification of the  
3148.2:2009 and to confirm that the ICP-OES  
technique meets AOAC criteria for calcium  
determination in animal feed. The results of  
this research are expected to provide  
theoretical benefits in the development of  
analytical science, as well as practical  
benefits for testing laboratories and the feed  
industry in ensuring the quality of their  
products.  
method's  
performance  
parameters  
according to SNI 3148.2:2009 guidelines,  
which have not been systematically studied  
in previous research. This standard also  
includes verification parameters such as  
linearity,  
detection  
accuracy,  
limit  
precision,  
and  
method  
limit of  
(LOD),  
Quantitation (LOQ), which serve as a  
reference for validating laboratory test  
results. Linearity parameter verification is  
performed to ensure that the instrument's  
emission signal is directly proportional to the  
calcium concentration within a specific  
range, while the limit of the linearity curve is  
needed to determine the maximum limit of  
the calibration curve [13]. Precision indicates  
the reproducibility of results under the same  
testing conditions [14], while accuracy  
describes the closeness between the test  
result values and the true value [15].  
Meanwhile, LOD and LOQ are important for  
knowing the minimum concentration limit  
Materials and Methods  
Materials  
The equipment used in this experiment  
consists of two types: main equipment and  
supplementary  
equipment.  
The  
main  
equipment used is the ICP-OES instrument  
(PerkinElmer Optima 8300) with a plasma  
power of 1400 W. Plasma gas flow 12 L/min,  
Auxiliary gas flow 1 L/min, Nebulizer gas flow  
0.8 L/min and axial view observation mode.  
Calibration was internal using a single-  
element calibration curve. Meanwhile, the  
315  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
supplementary equipment includes the  
Mettler Toledo analytical balance, furnace,  
100 mL and 200 mL volumetric flasks,  
micropipette, 10 mL measuring pipette,  
porcelain crucible, funnel, stirring rod,  
spatula, compressor, 0.45 μm filter paper, 50  
mL beaker, rubber bulb, and spray bottle.  
The materials used in the experiment consist  
of test materials and chemicals. The test  
materials include livestock feed samples and  
dissolve  
the  
remaining  
minerals.  
The  
solution was then cooled and filtered using  
0.45 μm filter paper. The obtained filtrate is  
placed into a 100 mL volumetric flask and  
diluted with distilled water to the mark.  
Preparation of Precision and Accuracy Test  
Solution  
The prepared test sample of 95 mL was  
placed into a 100 mL volumetric flask and 5  
mL of a 10 mg/L standard solution was  
added. For precision and accuracy testing,  
the test solution was prepared in ten  
replicates.  
a
standard  
calcium  
solution  
with  
a
concentration of 1000 mg/L. The chemicals  
used include concentrated nitric acid (HNO),  
HCl solution with concentrations of 0.5 M  
and 3 M, demineralized water (distilled  
water), and argon gas.  
Preparation of Limit of Detection Test  
Solution  
Preparation of 100 mg/L Calcium Standard  
Solution  
The instrument's detection limit test is  
conducted using a solvent blank. The  
preparation of the blank involves using 5 mL  
of concentrated HNO3, which is added to a  
100 mL flask that has previously been  
supplemented with a small amount of  
distilled water, then brought to the mark  
with distilled water and homogenized. The  
Limit of Detection test solution was made in  
ten replicates.  
A total of 10 mL of standard calcium solution  
with a concentration of 1000 mg/L was taken  
using a pipette, then placed into a 100 mL  
volumetric flask. Next, the volume of the  
solution was adjusted with distilled water  
until it reached the mark, then shaken until  
homogeneous.  
Preparation of Calcium Standard Series  
Solutions (0; 2; 4; 8; 16; 20) mg/L  
Preparation of Method Detection Limit and  
Limit of Quantitation Solutions  
5 mL of concentrated nitric acid (HNO) was  
pipetted into each 100 mL volumetric flask  
that contained a small amount of distilled  
water. After that, a standard calcium solution  
with a concentration of 100 mg/L was added  
to each volumetric flask in amounts of 0, 2,  
4, 8, 16, and 20 mL, respectively.  
A 20 mg/L standard solution of 5 mL was  
placed into a 100 mL volumetric flask, and  
adjusted to the mark with a sample of known  
concentration. The test solution for the  
method  
detection  
limit and Limit  
of  
Quantitationwas made in ten replicates.  
Test Sample Preparation  
Preparation of Confirmation Limit of  
Quantitation Solution  
A total of 2 g of feed sample was weighed  
and placed into a porcelain crucible, then  
burned in a furnace at 550 °C for 4 h until all  
the material turned to ash. The ash from the  
combustion was allowed to cool, then 10 mL  
of 3 M HCl solution was added, covered with  
a watch glass, and heated for 10 minutes to  
A total of 9 mL of standard solution with a  
concentration of 10 mg/L was added to a 100  
mL volumetric flask, then the volume of the  
solution was adjusted to the mark using a  
sample with a known concentration. The test  
316  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
solution  
for  
confirming  
the  
Limit  
of  
wavelength of 422.673 nm. The test results  
Quantitation was prepared in ten replicates.  
were  
statistically  
processed  
and  
then  
compared with the acceptance criteria set by  
AOAC. Data processing is calculated using  
Microsoft Excel.  
Linearity Test  
The standard working solution of Ca (0; 2; 4;  
8; 16; 20) mg/L that has been prepared was  
measured using the ICP-OES 422.673 nm  
instrument. The measurement results were  
Limit of Detection Test  
The prepared blank solution was analyzed  
using ICP-OES at a wavelength of 422.673  
nm. The obtained data were statistically  
processed and subsequently compared with  
the acceptance criteria established by AOAC.  
All data processing and statistical calculation  
were performed using Microsoft Excel.  
statistically  
processed  
to  
obtain  
the  
correlation value, slope, and intercept. The  
analysis results are then compared with the  
acceptance criteria set by AOAC. Data  
processing is calculated using Microsoft  
Excel.  
Limit Test of Linearity Curve (LoL)  
Method Detection Limit Test and Limit of  
Quantitation  
The working solution of the standard Ca with  
the lowest concentration of 2 mg/L and the  
highest concentration of 20 mg/L was  
measured ten times using the ICP-OES  
instrument at a wavelength of 422.673 nm.  
The measurement results were statistically  
processed to obtain the Fcalculated value.  
The prepared solutions for the Method  
Detection  
Limit  
(MDL)  
and  
Limit  
of  
Quantitation (LOQ) determination were  
analyzed using an Inductively Coupled  
Plasma-Optical Emission Spectrometer at a  
wavelength of 422.673 nm. The reading data  
The  
Fcalculated  
value  
obtained  
was  
were  
then  
statistically  
processed  
to  
compared with the acceptance criteria set by  
AOAC. Data processing is calculated using  
Microsoft Excel.  
determine the relative standard deviation  
(%RSD), percent recovery, signal-to-noise  
ratio (S/N), as well as the LDM and LOQ  
values. All evaluation results were compared  
with the acceptance criteria established by  
AOAC. All statistical calculation and data  
processing were calculated using Microsoft  
Excel.  
Precision Test (Repeatability)  
The precision test solution, which has been  
prepared ten times, was measured using the  
ICP-OES instrument at a wavelength of  
422.673  
statistically processed to determine the  
precision value or relative standard  
deviation (%RSD), and then compared with  
the acceptance criteria set by AOAC. Data  
processing is calculated using Microsoft  
Excel.  
nm.  
The  
test  
results  
were  
Confirmation Limit of Quantitation Test  
The confirmation solution for the Limit of  
Quantitationthat has been prepared was  
analyzed using the ICP-OES instrument at a  
wavelength  
of  
422.673  
nm.  
The  
measurement data were then statistically  
analyzed to calculate the relative standard  
deviation (%RSD) and recovery percentage.  
Accuracy Test  
These  
values  
are  
then  
evaluated  
by  
The accuracy test solution, which has been  
prepared ten times, was then measured  
comparing them against the acceptance  
limits set by AOAC. Data processing is  
calculated using Microsoft Excel.  
using  
the  
ICP-OES  
instrument  
at  
a
317  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
Result and Discussion  
the ground state to a higher energy level  
[18]. Based on the method verification tests  
that have been conducted, the results  
include linearity parameters, Linearity Curve  
Limit (LoL), Limit of Detection (LoD), Method  
Detection Limit (MDL), Limit of Quantitation  
(QL), Precision, and Accuracy. The results of  
the method verification test for determining  
calcium in animal feed can be seen in Table  
1.  
Method verification for calcium (Ca) content  
analysis in animal feed was conducted using  
the ICP-OES instrument. The basic principle  
of this technique is the measurement of the  
intensity of radiation emitted by atoms  
during energy transitions due to excitation  
[17]. ICP-OES operates based on the  
characteristic of atoms that absorb light  
energy and trigger electron excitation from  
Table 1. Data Results of Verification of the Calcium Content Determination Method in Animal  
Feed Using ICP-OES  
No  
Parameters  
Results  
Acceptance  
Conditions  
r ≥ 0.9950  
Source  
Conclusion  
1
Linearity (r)  
(0-20) mg/L  
r = 0.9991  
AOAC  
(2020)  
Meet The  
Requirements  
2
3
Limit Curve  
Linearity (2 and  
20) mg/L  
Precision  
(Repeatability)  
0.03 < 3.18  
Fcount < Ftable  
AOAC  
(2020)  
Meet The  
Requirements  
%RSD = 2.18  
%RSD ≤ ½  
CV Horwitz  
AOAC  
(2020)  
Meet The  
Requirements  
½ CV Horwitz =  
7.92  
(2.18≤ 7.92)  
%Recovery =  
(90-99)%  
Accuracy  
(Recovery)  
%Recovery =  
(80-115)%  
AOAC  
(2020)  
Meet The  
Requirements  
4
Limit of  
Detection (LoD)  
Method  
Detection Limit  
(MDL) and Limit  
of Quantitation  
(LoQ)  
0.0745 mg/L  
Positive  
Response  
Reads  
AOAC  
(2020)  
AOAC  
(2020)  
Meet The  
Requirements  
Meet The  
5
6
Theoretical  
MDL= 0.8079  
mg/L  
S/N = 9.08  
%Recovery =  
(86-113)%  
LoQ = 0.9060  
mg/L  
positive  
Requirements  
2.5-10  
%Recovery=  
(80-115)%  
Positive  
Response  
%RSD≤2/3  
CV  
Confirmation  
Limit of  
Quantitation  
%RSD = 10,  
2/3 CV Horwitz =  
10.2  
AOAC  
(2020)  
Meet The  
Requirements  
7
Horwitz  
10≤10.2  
Recovery = (86-  
113)%  
Recovery =  
(80-115)%  
318  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
Linearity  
correlation coefficient (r) value of 0.9991,  
which is close to 1. This indicates a strong  
Linearity describes the extent to which an  
analytical method can produce a response  
that is proportional to the concentration of  
the analyte within a certain range. This linear  
relationship is usually indicated by the  
correlation coefficient (r) [19]. The linearity  
calculation results are shown in Table 2, and  
the calibration curve for the Ca standard  
series obtained is shown in Figure 1.  
and  
proportional  
linear  
relationship  
between the concentration of the element  
Ca and the instrument response. The value  
of this correlation coefficient also meets the  
AOAC acceptance criteria, which is r ≥  
0.9950. Linearity evaluation was performed  
using six calibration points (0.0020.00  
mg/L), and the regression results showed  
excellent linearity for Ca determination  
using ICP-OES. The regression model yielded  
a coefficient of determination (R²) value of  
0.99842, indicating that over 99% of the  
variation in the optical signal can be  
explained by changes in Ca concentration.  
Based on the graph, the linearity test results  
show that the intensity of the calcium signal  
is directly proportional to the calcium  
concentration. The regression equation  
obtained is y = 293,320x 163,402 with a  
Table 2. The linearity calculation results  
Concentration  
Intensity  
No  
of Ca (mg/L)  
(Xi)  
(Xi)*(Yi)  
Xi2  
Yi2  
(Yi)  
0.0  
0
1
0.00  
0.00  
0.00  
0.00  
2
3
4
5
2.00  
4.00  
8.00  
16.0  
20.00  
343,348.52  
931,870.22  
2,124,684.91  
4,559,671.64  
5,726,023.29  
13,685,598.58  
686,697.04  
372,7480.88  
16,997,479.28  
72,954,746.24  
114,520,465.8  
208,886,869.2  
293,320.25  
4
16  
64  
256  
400  
740  
117,888,206,186.19  
868,382,106,922.85  
451,428,596,6781.71  
20,790,605,464,620.30  
32,787,342,717,622.40  
59,078,504,462,133.50  
6
Amount  
Slope (b)  
Intercept  
(a)  
50.00  
163,402.32  
r
0.9991  
0.9984  
R2  
(Sy/x)  
104,897.86  
In  
linear  
regression  
analysis,  
the  
between two datasets [21]. An ideal  
relationship occurs when the intercept value  
approaches zero and r approaches +1 or −1,  
relationship between variables is expressed  
through the equation y = bx + a, where b is  
the slope of the line, a is the intercept, x  
represents the analyte concentration, and y  
denotes the instrument response [20]. The  
correlation coefficient (r) is used to assess  
the strength of the linear relationship  
depending  
correlation.  
on  
the  
direction  
of  
the  
From the obtained curve, the intercept value  
(a) is -163,402.32. An intercept close to zero  
319  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
indicates that the instrument does not  
provide a signal when a blank solution is  
tested [22]. However, the presence of  
interference, noise, contamination, or other  
instrument effects often causes a signal to  
still be read even for a blank. A slope value  
(b) of 293,320.25 indicates the sensitivity  
level of the method. The larger the slope  
value, the higher the method's sensitivity to  
the analyte being tested [23].  
Figure 1. Curve of Calcium Concentration vs. Intensity  
Residual analysis showed random  
a
statistical parameters confirm that the  
calibration curve satisfies the linearity  
requirements recommended by AOAC and  
ISO/IEC 17025, demonstrating that the  
method is suitable for the determination of  
Ca in animal feed.  
distribution around the zero line with no  
observable systematic pattern, indicating  
absence of lack-of-fit and confirming that  
the linear model is adequate over the tested  
concentration range. The standard error of  
regression  
demonstrates  
relationship between analytical intensity  
(Sy/x)  
value  
of  
104,897.86  
Limit Curve Linearity  
good  
precision  
in the  
The linearity limit of the curve was measured  
by comparing with the standard solutions of  
Ca using concentrations 2 mg/L and 20 mg/L,  
each 10 times. The results of the two  
standard deviations obtained are compared.  
The data limits for the linearity curve and the  
and  
confidence  
analyte  
concentration.  
interval for  
The  
the  
95%  
slope  
(277,123.40309,517.10) indicates stable  
instrument sensitivity, while the confidence  
interval for the intercept (343,277.29 to  
16,472.63) suggests that the intercept does  
not significantly differ from zero, indicating  
the absence of systematic bias at the zero  
F-table  
are  
shown  
in  
Table  
3.  
The  
comparison of the standard deviation of the  
data from the repeated tests is in the form of  
F-count. The calculated F count is 0.018.  
Based on degrees of freedom df1=df2=n-1  
and a confidence level of 95% (α = 0.05), the  
value of F-tabel is obtained as 3.18. From the  
calculation data, it is known that F-count < F-  
concentration  
level.  
Although  
measurements were performed without  
replication at each concentration level, the  
calibration design remains appropriate for  
initial method verification. Overall, these  
320  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
table, so H0 is accepted, meaning the  
precision level of the lowest working solution  
intensity is not significantly different from  
the precision level of the highest working  
solution intensity [24]. Based on the results  
obtained, with a confidence level of 95% (α =  
0.05), 2 mg/L 20 mg/L is a linear regression  
and 20 mg/L is the highest concentration  
detectable by the instrument.  
Table 3. Data limits for linearity curve and F-Table  
Intensity  
Sample  
Lowest Standard  
Highest Standards  
(20 mg/L)  
(2 mg/L)  
1
2
3
4
5
6
7
8
350,226.9894  
332,253.1689  
349,590.3771  
350,098.3865  
347,700.5273  
332775,5831  
333,089.9671  
343,347.0374  
347,605.1231  
338,383.0410  
8,935.336  
5,773,543.8410  
5,741,227.0290  
5,697,047.2310  
5,579,630.9190  
5,658,376.0490  
5,652,233.3640  
5,689,108.3010  
5,658,277.0960  
5,625,120.0700  
5,624,015.7550  
51,372.977  
9
10  
SD  
SD2  
79,840,222.0  
2,639,182,775.4  
Fcount  
Ftable  
(df = n-1, α = 5%)  
0.018  
3.18  
Precision (Repeatability)  
objective comparison of method precision  
across different concentration levels, with  
lower values indicating higher consistency  
and analytical reliability. The experimental  
results yielded a %RSD of 2.18%, which was  
evaluated against the Horwitz coefficient of  
variation (CV) of 7.92%. The Horwitz CV,  
derived from an empirical model, serves as a  
widely accepted benchmark for assessing  
acceptable precision in chemical analysis.  
Precision testing is an essential parameter in  
analytical method validation, intended to  
evaluate the degree of agreement among  
repeated measurements obtained under  
identical analytical conditions. Precision  
reflects the magnitude of random errors in  
an analytical procedure and provides insight  
into the short-term reproducibility of the  
method. In this study, precision was  
assessed through a repeatability test, in  
which replicate analyses were performed  
using the same analyst, instrumentation,  
According  
established by AOAC, a method meets  
repeatability requirements when the  
to  
the  
acceptance  
criteria  
observed %RSD does not exceed 0.5 × CV  
Horwitz.  
reagents,  
and  
experimental  
conditions  
within a limited time frame [25].  
In this case, the obtained %RSD was  
significantly lower than the allowable limit  
(3.96%), confirming that the analytical  
The repeatability of the method was  
expressed as the relative standard deviation  
(%RSD),  
a
statistical  
parameter  
that  
method  
demonstrates  
satisfactory  
describes data dispersion relative to the  
mean value. The use of %RSD allows for  
repeatability. These results indicate that  
321  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
random  
variability  
is  
well  
controlled,  
RSD ≤ 1%  
: very accurate  
(1)  
(2)  
(3)  
(4)  
supporting the method’s reliability and  
suitability for routine quantitative analysis  
[26]. The relative standard deviation value  
obtained from the precision test indicates  
that the precision test results fall into the  
medium accuracy category.  
1% < RSD ≤ 2% : high accuracy  
2% < RSD ≤ 5% : moderate accuracy  
RSD ≥ 5 %  
: low accuracy  
Table 4. Precision test data and calculations  
C
C
V
V
C
Actual  
Test  
Sampel  
(mg/L)  
Spike  
(mg/L)  
Spike  
(mL)  
Sampel  
(mL)  
Target  
(mg/L)  
Intensity  
Concentration  
of Ca (mg/L)  
1.022381  
1.03481  
1
2
3
4
5
6
7
8
9
132,409.6  
136,044.6  
140,934.6  
145,760.1  
145,603.9  
149,725.7  
143,037.5  
151,478  
1.051531  
1.068031  
1.067497  
1.08159  
1.058721  
1.087582  
1.094452  
1.080174  
10.64677  
1.064677  
0.023  
0.6600  
10  
5
95  
1,1270  
153,487.3  
149,311.4  
10  
Amount  
Average  
SD  
% RSD  
0,5 CV Horwitz  
2.18  
7.92  
Accuracy (Recovery)  
using a standard solution with a known  
concentration. The results of the accuracy  
test and the corresponding calculations are  
presented in Table 5.  
Accuracy is a key parameter in analytical  
method  
closeness  
validation  
of agreement  
that  
describes  
between  
the  
the  
measured value and the true or accepted  
reference value of an analyte [27]. It reflects  
the trueness of an analytical method and  
indicates the presence of systematic error. In  
quantitative analysis, accuracy is commonly  
evaluated through recovery studies, in which  
a known amount of analyte standard is  
spiked into the sample matrix and analyzed  
using the proposed analytical method. This  
approach allows evaluation of matrix effects,  
analyte loss during sample preparation, and  
The obtained %Recovery values fall within  
the acceptable limits recommended by the  
AOAC guidelines for analytical method  
validation, indicating that the method  
produces reliable and unbiased results.  
Based on ten replicate analyses, the  
%Recovery values ranged from 90% to 99%,  
with an average recovery of 94%. These  
results comply with the acceptance criteria  
established by AOAC (2002), which specify an  
acceptable recovery range of 80115% for  
quantitative analytical methods. Therefore,  
the method.  
potential  
analytical  
bias.  
Accuracy  
is  
generally expressed as the percentage  
recovery (%Recovery) of the added analyte  
322  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
Table 5. Accuracy Data and Calculation  
Actual  
Concentration  
C
C
V
V
C
Test  
Sampel  
(mg/L)  
Spike  
(mg/L)  
Spike  
(mL)  
Sampel  
(mL)  
Target  
(mg/L)  
Intensity  
% Recovery  
Calcium  
(mg/L)  
1.0224  
1.0348  
1.1045  
1.0515  
1.1163  
1.0675  
1.0945  
1.0587  
1.0008  
1.1258  
1.0677  
1
2
3
4
5
6
7
8
132,409.6  
136,044.6  
156,432.9  
140,934.6  
159,879.9  
145,603.9  
153,487.3  
143,037.5  
126,101.8  
162,660.6  
90  
91  
98  
93  
99  
94  
97  
93  
88  
99  
94  
0.6600  
10  
5
95  
1.1270  
9
10  
Average  
Accuracy Range Value  
Acceptance Conditions  
(90-99) %  
(80-115) %  
Limit of Detection (LoD)  
parameter reflects the minimum sensitivity  
of an analytical method. The LoD can be  
determined by measuring blank samples,  
which are matrices free of the analyte, to  
confirm that the instrument response  
originates from the presence of the analyte  
rather than background noise [29].  
The data and LoD calculations are presented  
in Table 6. The limit of detection is defined as  
the lowest concentration of an analyte in a  
sample that produces a signal significantly  
different from that of the blank [28]. This  
Table 6. Data and LoD Calculations  
2
Test  
1
2
3
4
5
6
7
Intensity  
17,400.6287  
16,400.4184  
3,920.1495  
821.1895  
1,259.3401  
2,038.0774  
338.3124  
285.8692  
235.6276  
12,393.7679  
3.0223  
Concentration (xi) (mg/L)  
(xi-  
 
(xi-  
 )  
0.3443  
0.3408  
0.2966  
0.2856  
0.2872  
0.2899  
0.2839  
0.2837  
0.2836  
0.3022  
0.04207  
0.03853  
-0.00562  
-0.01658  
-0.01503  
-0.01228  
-0.01829  
-0.01848  
-0.01866  
0.02435  
0.0018  
0.0015  
0.0001  
0.0003  
0.0002  
0.0002  
0.0003  
0.0003  
0.0003  
0.0006  
0.0056  
8
9
10  
Amount  
Average (mg/L)  
SD  
0.3022  
0.0248  
LoD (mg/L)  
0.0745  
323  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
In this study, the theoretical LoD value  
obtained was 0.0745 mg/L, representing the  
lowest concentration of Ca that could be  
reliably detected by the instrument. The LoD  
evaluation met the acceptance criteria  
established by AOAC (2002), as it produced a  
sample  
that can  
still  
be  
determined  
quantitatively with acceptable levels of  
precision and accuracy, under agreed-upon  
test conditions [31]. The LOQ value and  
method detection limit were determined by  
measuring the prepared solutions using the  
analyte addition (spiking) technique. Data  
and Calculations for the Detection Limit of  
the Method and Limit of Quantitationare  
shown in Table 7.  
positive  
and  
distinguishable  
analytical  
response  
Method Detection Limit  
Determining the method's detection limit  
reflects both the laboratory's capabilities  
and limitations in applying an analytical  
In this test, the theoretical concentration  
values for MDL and LoQ were obtained. The  
method's limit of detection is acceptable if  
the data from the test repetitions meet  
several acceptances following limit formula  
5-9.  
method  
to  
detect  
analytes  
at  
low  
concentrations. The method's limit of  
detection is defined as the concentration of  
an analyte that can be detected thru all  
stages of the analysis procedure with 99%  
confidence, where the signal produced can  
be clearly distinguished from the blank  
signal [30]. Limit of Quantitation (LOQ), often  
referred to as the reporting limit, is the  
lowest concentration of an analyte in a  
%RSD < 0,67 Horwitz  
(5)  
(6)  
(7)  
(8)  
(9)  
% Recovery = 80-115 % (AOAC, 2002)  
The signal to noise ratio (S/N) = 2,5-10  
MDL < Spike < 10MDL  
LoQ ≤ BML  
Table 7. Data and Calculation of Method Detection Limit and Limit of Quantitation  
C
C
V
V
C
C
Recovery  
(%)  
Test  
Sampel  
(mg/L)  
Spike  
(mg/L)  
Sampel  
(mL)  
Spike  
(mL)  
Target  
(mg/L)  
Intensity  
Calcium  
(mg/L)  
2.9500  
2.8900  
2.3000  
2.2800  
2.6000  
2.6500  
2.2500  
2.8900  
2.8600  
2.3500  
26.0200  
2.6020  
0.2864  
0.0906  
9.08  
1
9,762,972 ,568  
1,006,484,546  
1,047,067,863  
1,092,602,128  
1,129,398,123  
1,125,396,591  
1,150,724,794  
1,179,239,004  
1,554,535,717  
2,011,566,989  
113 %  
111%  
88%  
2
3
4
87%  
5
100%  
102%  
86%  
0.6600  
20  
90  
10  
2.5940  
6
7
8
111%  
110%  
90%  
9
10  
Amount  
Average  
SD  
SD’ SD /√  
S/N  
MDL theoretical (mg/L)  
10 MDL (mg/L)  
0.8079  
8.079  
LoQ theoretical (mg/L)  
0.906  
The method detection limit (MDL) must  
satisfy acceptable precision and accuracy  
criteria. Precision is commonly expressed as  
the coefficient of variation, also referred to  
as the relative standard deviation (%RSD).  
The  
%RSD  
obtained  
from  
repeated  
324  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
measurements  
process should not exceed 0.67 times the  
Horwitz coefficient of variation (CV) value  
during  
the  
verification  
baseline method limit (BML) with the  
corresponding limit of quantitation (LoQ).  
This approach is based on the established  
working range of the method, in which the  
lower limit is defined by the LoQ and the  
upper limit by the limit of linearity (LoL).  
Consequently, the LoQ serves as a critical  
parameter for determining the applicability  
and reliability of an analytical method for  
quantitative measurements [34].  
[32].  
In  
addition,  
the  
recovery  
test  
(%Recovery) results must fall within the  
acceptable range of 80115%.  
The signal-to-noise ratio (S/N) is used to  
assess the level of random error that may  
occur during the testing process and to  
estimate the precision of repeated test  
results which is calculated using Microsoft  
Excel. If the S/N ratio is less than 2.5, this  
indicates that the random variation in the  
replicates is quite high and may cause the  
method detection limit (MDL) to be large. In  
such conditions, the concentration of the  
added analyte (spike) needs to be increased  
to produce a stronger signal. Conversely, if  
the S/N exceeds 10, it means the amount of  
Based on the limit of quantitation test  
results, the theoretical LoQ was determined  
to be 0.9060 mg/L, representing the lowest  
analyte concentration that can be quantified  
with acceptable precision and accuracy. This  
theoretical value required confirmation to  
ensure  
compliance  
with  
method  
performance requirements. Therefore, a  
LoQ confirmation test was conducted. The  
results demonstrated that the method met  
the required precision and accuracy criteria  
and complied with the acceptance standards  
established by the AOAC (2002), confirming  
the suitability of the method for quantitative  
analysis  
analyte  
added  
is  
too  
high,  
so  
its  
concentration needs to be reduced to  
remain within the ideal working range [33].  
When determining the method detection  
limit (MDL), the choice of spike concentration  
must be made carefully to ensure the results  
obtained are within an acceptable range. The  
precision level in determining the MDL is  
highly influenced by the concentration of the  
spike used. If the obtained MDL value  
exceeds the spike level, it will be statistically  
difficult to distinguish between the spike  
value and the blank value, resulting in low  
precision. Therefore, the determination of  
spike levels should consider the lower limit  
of the concentration range being verified, as  
well as follow the recommended ratio, which  
is MDL : LoQ = 4 : 10.  
Confirmation Limit of Quantitation (LoQ)  
The  
data  
and  
calculations  
for  
the  
confirmation of the limit of quantitation  
(LoQ) are presented in Table 8. Based on the  
quantitation test results, the theoretical LoQ  
was determined to be 0.9000 mg/L, which  
represents the lowest analyte concentration  
that can be quantified reliably and therefore  
requires confirmation to ensure that the  
method meets the required precision and  
accuracy criteria [35]. LoQ confirmation was  
carried out by analyzing a calcium standard  
solution with a concentration of 10 mg/L  
When the method detection limit (MDL)  
meets the required statistical acceptance  
using  
ICP-OES.  
The  
precision  
of  
the  
criteria,  
the  
obtained  
value  
may  
be  
measurements was evaluated in terms of  
the relative standard deviation (%RSD), and  
the confirmation test yielded an %RSD value  
of 10%. This value satisfies the applicable  
acceptance criterion, namely %RSD ≤ 2/3 CV  
compared with relevant environmental  
laboratory quality standards. In routine  
practice, laboratories that have verified their  
analytical methods may also compare the  
325  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
Horwitz,  
indicating  
that  
the  
analytical  
LoQ is suitable for reliable quantitative  
method demonstrates adequate precision at  
the quantitation limit and that the confirmed  
determination  
concentration levels.  
of  
calcium  
at  
low  
Table 8. Data and Calculation for Confirmation Limit of Quantitation  
Ca  
Sample  
(mg/L)  
C
C Ca  
actual  
(mg/L)  
1.3014  
1.2876  
1.3252  
1.3546  
1.3541  
1.3795  
1.3196  
1.2821  
1.7114  
1.5635  
13.8790  
1.3879  
0.1395  
10.0  
C
Vc  
(mL)  
Vd  
(mL)  
Recovery  
Test  
Target  
(mg/L)  
Intensity  
xi-x  
̅
(xi-x̅)2  
(mg/L)  
(%)  
1
2
3
4
5
6
7
8
371,290  
367,644  
377,557  
385,331  
385,191  
391,908  
376,094  
366,178  
479,541  
440,481  
86  
85  
88  
90  
90  
91  
87  
85  
114  
104  
-0,0864  
-0,1002  
-0,0627  
-0,0333  
-0,0338  
-0,0083  
-0,0682  
0,1058  
0,3234  
0,1755  
0.0075  
0.0101  
0.0039  
0.0011  
0.0011  
0.0001  
0.0047  
0.0112  
0.1046  
0.308  
0.6600  
10  
91  
9
1.5006  
9
10  
Amount  
Average  
SD  
% RSD  
% Recovery  
2/3 CV Horwitz  
(85-114) %  
10.2  
Conclusions  
animal feed, both for quality control and  
regulatory compliance purposes. However,  
this study has certain limitations, particularly  
The verification results demonstrate that the  
ICP-OES method based on SNI 3148.2:2009  
provides adequate analytical performance  
for the determination of calcium in livestock  
the  
absence  
of  
a
comprehensive  
evaluation.  
measurement  
uncertainty  
Therefore, further research is recommended  
to assess the influence of different feed  
feed  
matrices.  
The  
method  
exhibited  
excellent linearity, stable precision, and  
satisfactory accuracy within the acceptance  
limits recommended by the AOAC, indicating  
a consistent and proportional instrument  
matrices  
and  
to  
perform  
structured  
measurement uncertainty calculations in  
accordance with ISO/IEC 17025. Addressing  
these aspects would further strengthen  
method validity and broaden its application  
in industrial-scale laboratory testing.  
response  
to  
variations  
in  
calcium  
concentration. Furthermore, the low values  
of the limit of detection (LOD), limit of  
quantitation (LoQ), and method detection  
limit (MDL) confirm that the method is  
sufficiently sensitive to detect and quantify  
Acknowledgement  
This research supported by the Ministry of  
Industry Affairs Republic Indonesia through  
collaboration research by lecturers from  
Polytechnic ATI Padang.  
calcium  
at  
low  
concentration  
levels,  
supporting its applicability for feed quality  
testing across diverse matrix compositions.  
Overall, the verification confirms that ICP-  
OES is a reliable and suitable analytical  
technique for routine calcium analysis in  
Author Contributions  
326  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
The contribution of each author to this  
article is : "Conceptualization, Syafrinal and  
Nurmaliza.; Methodology, Syafrinal and  
2023.  
391446.  
Nurmaliza;  
Software,  
Hafnimardiyanti.;  
[4]. Wu, G. Management of metabolic  
Validation, Syafrinal., Nurmaliza and Selfa  
Dewati Samah.; Formal Analysis, Pevi Riani.;  
Investigation, Renny Futeri.; Resources,  
Syafrinal.; Data Curation, Nurmaliza.; Writing  
Original Draft Preparation, Syafrinal and  
Nurmaliza; Writing Review & Editing,  
Syafrinal.; Visualization, Hafnimardiyanti.;  
disorders  
diseases)  
ruminant  
Agriculture.  
(including  
ruminant  
animals.  
2020.  
metabolic  
in  
and  
In  
non-  
Animal  
471491).  
[5]. Carter, S., Clough, R., Fisher, A., Gibson,  
B., Russell, B., & Waack, J. Atomic  
Supervision,  
Administration,  
Pevi  
Selfa  
Riani.;  
Project  
Dewati  
Samah.;  
spectrometry  
update:  
Review  
of  
Funding Acquisition, Renny Futeri.  
advances in the analysis of metals,  
chemicals and materials. Journal of  
Analytical Atomic Spectrometry, 34(11):  
Conflict of Interest  
21592216.  
2019.  
The authors declare that there is no conflict  
of interest regarding the publication of this  
paper.  
[6]. Khan, S. R., Sharma, B., Chawla, P. A., &  
Bhatia, R. Inductively coupled plasma  
optical emission spectrometry (ICP-  
OES): A powerful analytical technique  
for elemental analysis. Food Analytical  
Ethical Standards  
This article does not contain any studies  
involving human or animal subjects.  
Methods,  
2022.  
15(3):  
666688.  
References  
[7]. Zhang, J. The role of verification and  
validation process in best estimate plus  
[1]. Moorby, J. M., & Fraser, M. D. New  
feeds and new feeding systems in  
intensive and semi-intensive forage-  
fed ruminant livestock systems: A  
uncertainty  
development. Nuclear Engineering and  
Design, 355: 110312. 2019.  
methodology  
review.  
Animal,  
2021.  
15:  
[8]. Oreste, E. Q., et al. Evaluation of  
sample preparation methods for the  
determination of Ca, Cu, Fe, K, and Na  
in milk powder samples by ICP-OES.  
Food Analytical Methods, 2016. 9(3):  
777784.  
[9]. Phan-Thien, K.-Y., Wright, G. C., & Lee,  
N. A. Determination of essential  
minerals in peanuts by ICP-MS and ICP-  
OES. Food Chemistry, 2012. 134(1): 453–  
460.  
[2]. Sonea,  
C.,  
Gheorghe-Irimia,  
R.-A.,  
Tăpăloagă, D., & Tăpăloagă, P.-R.  
Nutrition and animal agriculture in the  
21st century: A review of future  
prospects. Annals of the University of  
Craiova  
Agriculture, Montanology,  
Cadastre Series, 2023. 53(1): 303312.  
[3]. Malik,  
D.,  
Narayanasamy,  
N.,  
Pratyusha, V. A., Thakur, J., & Sinha, N.  
Inorganic nutrients: Macrominerals. In  
Textbook of Nutritional Biochemistry,  
327  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
[10]. Kargarghomsheh, P., et al. Evaluation  
of elements in Persian tahini using ICP-  
[16]. Riviere, J. E. Chemistry and limits of  
analytical detection and quantitation.  
In Zero Much to Do About Nothing? (pp.  
[17]. Aldakheel, R. K., Gondal, M. A., Nasr, M.  
M., Almessiere, M. A., & Idris, N.  
Spectral analysis of Moringa oleifera  
leaves using XPS, LIBS and ICP-OES.  
Talanta, 2020. 217: 121062. 2020.  
OES.  
Environmental  
2025. 105(8):  
International  
Analytical  
17231737.  
Journal  
Chemistry,  
2025.  
of  
[11]. Düzgün, M., & Kahraman, Ş. Evaluation  
of mineral safety index, nutritional and  
risk assessment of sorghum genotypes  
using  
ICP-OES.  
Journal  
of  
Food  
[18]. Rehan, I., Gondal, M. A., Rehan, K., &  
Sultana, S. Spectral diagnosis of toxic  
elements in face powders using LIBS  
and ICP-OES. Talanta, 2020. 217:  
121007.  
Composition and Analysis, 2025. 146:  
107960.  
[12]. Betz, J. M., Brown, P. N., & Roman, M. C.  
Accuracy, precision, and reliability of  
chemical measurements in natural  
products research. Fitoterapia, 2010.  
[19]. Jarantow, S. W., Pisors, E. D., & Chiu, M.  
L. Linear and nonlinear regression  
82(1):  
4452.  
2011.  
analysis  
in  
quantitative  
biological  
assays. Current Protocols, 2023. 3(6).  
[20]. Stalikas, C., & Sakkas, V. Calibration and  
[13]. Jurado, J. M., Alcázar, A., Muñiz-  
Valencia, R., Ceballos-Magaña, S. G., &  
Raposo, F. Practical considerations for  
linearity assessment of calibration  
curves. Talanta, 2017. 172: 221229.  
2017.  
correlation  
Practical considerations. Microchimica  
Acta, 2024. 191(2): 81.  
in  
chemical  
analysis:  
[14]. McAlinden, C., Khadka, J., & Pesudovs,  
[21]. Hazra, A., & Gogtay, N. Correlation and  
linear regression. Indian Journal of  
Dermatology, 2016. 61(6): 593. 2016.  
K.  
Precision  
(repeatability  
and  
reproducibility) studies and sample-  
size calculation. Journal of Cataract and  
Refractive Surgery, 2015. 41(12): 2598–  
[22]. Brunetti, D. E. About estimating the  
limit of detection by the signal-to-noise  
approach. Pharmaceutical Analytical  
2604.  
2015.  
[15]. Mehl, A., Reich, S., Beuer, F., & Güth, J.  
F. Accuracy, trueness, and precision: A  
Acta,  
2015.  
6(4).  
2015.  
[23]. González, A. G., & Herrador, M. Á.  
Practical guide to analytical method  
validation. TrAC Trends in Analytical  
guideline  
for  
digital  
dentistry.  
International Journal of Computerized  
Dentistry, 2021. 24(4): 341352.  
Chemistry,  
2007.  
26(3):  
227238.  
328  
Syafrinal., et al.  
Chempublish Journal, 9(2) 2025, 313-329  
[24]. Lestari, S., Watini, S., & Rose, D. E.  
Impact of self-efficacy and work  
discipline on employee performance.  
[30]. Gegenschatz, S. A., Chiappini, F. A.,  
Teglia, C. M., Muñoz de la Peña, A., &  
Goicoechea, H. C. Computing limits of  
detection  
univariate calibration. 2022. Analytica  
Chimica Acta, 1209: 339342.  
and  
quantification  
in  
Aptisi  
Transactions  
on  
Technopreneurship, 2024. 6(2): 270–  
[31]. Khamis, M. M., Adamko, D. J., & El-  
Aneed, A. Method development and  
validation for absolute quantification  
[25]. Prenesti, E., & Gosmaro, F. Trueness,  
precision and accuracy: A critical  
overview. Accreditation and Quality  
Assurance, 2015. 20(1): 3340. 2015.  
[26]. Nair, A., Chandrashekhar, H. R., &  
Nayak, U. Y. RP-HPLC method for  
quantification of cefotaxime sodium  
of endogenous metabolites.  
Mass  
Spectrometry Reviews, 2021. 40(1): 31–  
[32]. Ehling, S., Thompson, J. J., Schimpf, K. J.,  
Pacquette, L. H., & Haselberger, P. A.  
Precision  
of  
modern  
analytical  
methods in food analysis. Journal of  
using  
Journal  
Science.  
green  
of  
analytical  
approach.  
AOAC International, 108(4): 566571.  
2025.  
Applied  
Pharmaceutical  
2024.  
[33]. Peters, F. T.,  
Bioanalytical method validation in  
forensic and clinical toxicology.  
Accreditation and Quality Assurance,  
7(11): 441449. 2002.  
&
Maurer, H. H.  
[27]. Feinberg, M. Validation of analytical  
methods based on accuracy profiles.  
Journal of Chromatography A, 2007.  
1158(12):  
174183.  
[28]. Taleuzzaman, M. Limit of blank, limit of  
detection and limit of quantification.  
[34]. Uhrovčík, J. Strategy for determination  
of LOD and LOQ values. Talanta, 119:  
Organic  
International  
&
Medicinal  
Journal,  
Chemistry  
2018. 7(5).  
178180.  
2014.  
[35]. Peris-Vicente, J., Esteve-Romero, J., &  
Carda-Broch, S. Validation of analytical  
methods based on chromatographic  
techniques. In Analytical Separation  
[29]. Bulska, E., & Ruszczyńska, A. Analytical  
techniques  
determination.  
Reviews,  
for  
trace  
Physical  
2017.  
element  
Sciences  
2(5).  
Science,  
2015.  
5,  
17571808).  
329