Syafrinal., et al.
measurements
process should not exceed 0.67 times the
Horwitz coefficient of variation (CV) value
during
the
verification
baseline method limit (BML) with the
corresponding limit of quantitation (LoQ).
This approach is based on the established
working range of the method, in which the
lower limit is defined by the LoQ and the
upper limit by the limit of linearity (LoL).
Consequently, the LoQ serves as a critical
parameter for determining the applicability
and reliability of an analytical method for
quantitative measurements [34].
[32].
In
addition,
the
recovery
test
(%Recovery) results must fall within the
acceptable range of 80–115%.
The signal-to-noise ratio (S/N) is used to
assess the level of random error that may
occur during the testing process and to
estimate the precision of repeated test
results which is calculated using Microsoft
Excel. If the S/N ratio is less than 2.5, this
indicates that the random variation in the
replicates is quite high and may cause the
method detection limit (MDL) to be large. In
such conditions, the concentration of the
added analyte (spike) needs to be increased
to produce a stronger signal. Conversely, if
the S/N exceeds 10, it means the amount of
Based on the limit of quantitation test
results, the theoretical LoQ was determined
to be 0.9060 mg/L, representing the lowest
analyte concentration that can be quantified
with acceptable precision and accuracy. This
theoretical value required confirmation to
ensure
compliance
with
method
performance requirements. Therefore, a
LoQ confirmation test was conducted. The
results demonstrated that the method met
the required precision and accuracy criteria
and complied with the acceptance standards
established by the AOAC (2002), confirming
the suitability of the method for quantitative
analysis
analyte
added
is
too
high,
so
its
concentration needs to be reduced to
remain within the ideal working range [33].
When determining the method detection
limit (MDL), the choice of spike concentration
must be made carefully to ensure the results
obtained are within an acceptable range. The
precision level in determining the MDL is
highly influenced by the concentration of the
spike used. If the obtained MDL value
exceeds the spike level, it will be statistically
difficult to distinguish between the spike
value and the blank value, resulting in low
precision. Therefore, the determination of
spike levels should consider the lower limit
of the concentration range being verified, as
well as follow the recommended ratio, which
is MDL : LoQ = 4 : 10.
Confirmation Limit of Quantitation (LoQ)
The
data
and
calculations
for
the
confirmation of the limit of quantitation
(LoQ) are presented in Table 8. Based on the
quantitation test results, the theoretical LoQ
was determined to be 0.9000 mg/L, which
represents the lowest analyte concentration
that can be quantified reliably and therefore
requires confirmation to ensure that the
method meets the required precision and
accuracy criteria [35]. LoQ confirmation was
carried out by analyzing a calcium standard
solution with a concentration of 10 mg/L
When the method detection limit (MDL)
meets the required statistical acceptance
using
ICP-OES.
The
precision
of
the
criteria,
the
obtained
value
may
be
measurements was evaluated in terms of
the relative standard deviation (%RSD), and
the confirmation test yielded an %RSD value
of 10%. This value satisfies the applicable
acceptance criterion, namely %RSD ≤ 2/3 CV
compared with relevant environmental
laboratory quality standards. In routine
practice, laboratories that have verified their
analytical methods may also compare the
325